Description
OUTPUT CHARACTERSTICS
MODES | MAX. P-P VOLTAGE | RATED LOAD OHMS | MAX/ POWER (WATTS) | |
MONOPOLAR CUT |
PURE | 900 | 500 | 400 |
BLEND I | 1100 | 500 | 300 | |
BLEND II | 1200 | 500 | 250 | |
BLEND III | 1300 | 500 | 200 | |
BLEND IV | 1300 | 500 | 150 | |
MONOPOLAR COAG. |
SOFT | 1600 | 500 | 250 |
DESSICATE | 1800 | 500 | 200 | |
FORCE | 2100 | 500 | 150 | |
SPRAY | 3000 | 500 | 100 | |
BIPOLAR |
MICRO | 320 | 100 | 90 |
STANDARD | 460 | 100 | 90 | |
MICRO/CUT | 600 | 100 | 90 |
PPQM System
- Measuring Frequency : 85 kHz ± 10 kHz Measuring current : < 10μA
- Dual Area PREM Patient Return Electrode : 5-135Ω
- Single Area Patient Return Electrode : < 10ΩApplications FieldsPlastic, Neuro, ENT, Spine, Opthalmic, Laparo Surgery, Endo, GI, Ortho, Open & General Surgery, Onco, Gynecology, Hernia, GI Endocut, Papillotomy, Polypectomy, Under Water, Urology, Gynec TCRE, Hysteroscopy, Thoracotomy, Unit is suitable for all kind of Open, Endoscopic, Laparoscopic and Minimally Invasive Surgery Thoracic and Cardiac Surgery.Safety Standards
- Classification according to the EC- Directive 93/ 42/ EEC: ClassIIb
- General Safety Standards: IEC 60601-1-1, IEC60601-2-2
- EMI/EMC Compatibility Standard: IEC60601-1-2
- IS Safety Standard: IS7583
Leakage Current
- HFLeakage : Monopolar <150 mA RMS- Bipolar <60 mA RMS
- Lowfrequency : Normal polarity, intact ground : < 10 μA
- Normal polarity, ground open : < 50μA
INPUT POWER
Operating Voltage : 190-260 Volt 50Hz AC, (Optional 90-120Volts 60Hz AC)
DIMENSIONS
L x W x H – 350 x 350 x 130mm Net Weight in 5Kg
Reviews
There are no reviews yet.